Issue |
Date |
Cost |
FR590 installation |
1997 |
|
FR590 Generator on loan from Dep. Biostructures |
1997 |
|
SGI-02 Workstation etc. |
Jan 1997 |
$22,000 |
First X-ray tube |
Feb 1997 |
$3,891 |
Mounting Microscope |
Nov 1998 |
$3,100 |
Tube chiller line leak inside device |
1998 |
|
Generator Front Panel Replacement |
1999 |
$510 |
Mach3 controller signal tube check relais replaced |
1999 |
|
Server Upgrade to Win98 |
May 2000 |
$3000 |
Tube replacement |
2002 |
$2,700 ca. |
Donation of emmergency heat exchangers, pump |
2002 |
$200 |
Camera Chiller failure (starter ensemble) |
2003 |
|
Thermostate failure, required pt100 replacement |
2004 |
|
Reversed flush tube sieve cleaning |
2004 |
|
Exchanged Zero-Power Ram |
Jul 2005 |
|
Tube Chiller repair |
2007 |
|
Tube chiller replac2008ement |
2008 |
$6.440 |
Reversed flush tube sieve cleaning |
2008 |
|
Camera chiller repair (starter ensemble) |
2008 |
|
Camera Controller ADC change |
Jun 2008 |
$1,400 ca. |
Cappa CCD Camera Controller repair (water damage) |
Aug 2009 |
$800 |
Reversed flush tube sieve cleaning |
2012 |
|
Microscope camera replacement |
Apr 2012 |
$312 |
Pump replacement for R100 |
Dec 2013 |
$223 |
Reversed flush tube sieve cleaning |
2013 |
|
Purchase Signal Bulbs, Newark, 24V 1.2W (50 mA) |
Apr 2014 |
$115 (100 bulbs) |
Replacement of Server computer |
Sep 2015 |
$345 |
Reversed flush tube sieve cleaning |
2016 |
|
Camera chiller repair (new Fan) |
2016 |
|
Oxford Cryostream donation |
2017 |
|
Configured Firewall to protect Collect software |
Oct 2019 |
|
Linux-Client computer replaced |
Dec 2019 |
|
Reversed flush tube sieve cleaning |
2020 |
|
Added protective low temp thermostate to tube chiller |
Nov 2020 |
$14 |
Tube chiller pump coupling replacement |
Dec 2020 |
|
Cryostream and Dewar pump down |
July 2021 |
|
Mach3 Servo chip (IC 29, LM629N-8) replacement |
Mar 2021 |
|
AD41 compressor valve piston replacement |
Jun 2021 |
|
Purchased AD41 Compressor service kit |
Jun 2021 |
$280 |
Kappa CCD pump down (& adapter construct) |
Dez 2021 |
$100 (ca.) |
Tube Chiller and Tube sieve cleaned |
Nov 2022 |
|
AD41 compressor failure (loose cable) |
Nov 2022 |
|
Vacuum restored on 60 l lN2 Deware and Cryostream |
Nov 2022 |
|
Camera chiller 3-in-one starter assembly exchanged |
July 2023 |
$20 |
Princeton Instruments camera controller unit exchanged with one from ebay |
May 2023 |
$500 |
Vacuum restored on Cryostream |
May 2024 |
|